JC-3000F Pipe Flange Ultrasonic Flowmeter

The new generation JC3000F is developed, with the experience of 12 years of manufacturing ultrasonic flowmeter, based on our version-12 and version-15 flow meters.

  • Price Term: FOB/CIF/EXW/DAP
  • Package: Exported Wooden case or according to customer request
  • Delivery Time:  7 to 10 working days
  • Transports: By sea, by air, by high express
  • Shipment Port: Shanghai Port
  • Origin of Place: Jiangsu
  • Warranty: 12 Months

The new generation JC-3000F is developed, with the experience of 12 years of manufacturing ultrasonic flowmeter, based on our version-12 and version-15 flow meters. By using the high-performance MSP430FG4618 microprocessor made by Taxes Instruments, it is intended to provide our customers with lower cost but of high-performance general purpose ultrasonic flow meter.

Feature

◆ High Accuracy Measurement

Linearity: 0.5%, Repeatability: 0.2%, Accuracy: ±1%

◆ Easy Operation

It is carried out the menu operation by the magnetic bar in the main unit, or by four press button through opening the cover.

◆ Large measuring range

Several type transducers for selection, measuring pipe size is from DN15mm to  DN6000mm

◆ Low voltage and long life

Adopt low voltage, multi-pulse technology to improve accuracy, useful life, and reliability.

◆ Powerful Recording Function

record the totalizer data of the last 512  days/128 months/10 years. Automatic recording the power -on and power-off time and flow rate for the last 64 times, Automatic recording  the totalizer data  state for the last 32 days

◆ Long transmission distance, and easy network deployment

RS485 communication protocol, the ordinary cable can be applied

◆ Support heat energy measurement

Equipped with a temperature sensor, it can support heat energy measurement

 

Items Specifications
Main unit Accuracy Better than ±1%
Repeatability Better than 0.2%
Principle Transit-time measuring principle
Measurement

Period

500ms
Display LCD with backlight, display accumulated flow/heat, instantaneous flow/heat, velocity, time etc.
Output Analogue output: 4-20mA or 0-20mA current output. Impedance 0~1kW. Accuracy 0.1%.
OCT output: Frequency signal (1~9999HZ)
Relay output: over 20 source signal (no signal, reverse flow etc.)
RS485 serial port
Input Three analogue input
Three-wire PT100 resistor input (optional)
Other functions Automatically record the totaliser data of the last 64 days / 64 months / 5 years;

The power-on time and corresponding flow rate of the last 64 power on and off events. Allow manual or automatic flow loss compensation

The instrument working status of the last 64 days

pipe material Steel, stainless steel, cast iron, cement pipe, copper, PVC, aluminum, FRP etc. Liner is allowed
Size 15-6000mm
Straight pipe section In the upstream it must be beyond 10D, in the downstream it must be beyond 5D, in the upstream the length must be beyond 30D from the access of the pump. (D stands for pipe diameter)
Liquid Types Water, sea water, industrial sewage, acid & alkali liquid, alcohol, beer, all kinds of oils which can transmit ultrasonic single uniform liquid
Temperature Standard: -30˚C – 90˚C ,High-temperature:-30˚C – 160˚C
Turbidity Less than 10000ppm, with a little bubble
Flow Direction Bi-directional measuring, net flow/heat measuring
Environment Temperature Main Unit: -30˚C – 80˚C
Transducer: -40℃ -110℃, Temperature transducer: select on enquiry
Humidity Main Unit: 85% RH
Transducer: water-immersible, water depth less than 3m
Cable Twisted Pair Line, standard length of 20m, can be extended to 500m (not recommended); Contact the manufacturer for longer cable requirement.

RS-485 interface, transmission distance up to 1000m

Power Supply DC24V
Power Consumption Less than 1.5W
Protocols MODBUS, M-BUS, Fuji extended protocol and other factory protocol